High Precision PDF Data Logger EBI 310 | Type: EBI 310 TX
CHF 400.40*
Content:
1
Delivery according to order confirmation
Product number:
7643731
Manufacturer:
Xylem Analytics Germany (EBRO)
Length:
161 mm
Width:
118 mm
Height:
37 mm
Weight:
0.047 kg
* Specifications refer to the packaging unit
PDF data logger with different temperature sensors. Compared to the EBI 300, the EBI 310 is characterized by shorter measuring intervals and higher storage capacity.
EBI 310: Data logger with internal temperature sensor
EBI 310 TE: Data logger with external precision temperature sensor, internal temperature sensor can be used additionally. Measurements in high and low temperature range
EBI 310 TH: Data logger with external humidity and temperature sensor, internal temperature sensor can be used additionally. Monitoring the relative humidity in storage and transport
EBI 310 TX: Data logger with temperature dual adapter for two exchangeable temperature sensors (not included in the scope of supply), internal temperature sensor can be used additionally
- Automatic pdf-report creation with temperature/time values
- Transport of drugs, inoculants, blood products and primary products
- Transport of all temperature sensitive products
- Storage of reference samples in refrigerators and deep-freezers
- "Traffic light principle" by limit exceeded
- IQ/OQ procedures through standardized software and standard reporting process possible
- GMP, pharmaceutical law compliant; EN 12830, ATP, VO (EG) 37/205 compliant, 21 CFR Part 11 conform (Winlog.basic)
- Programmable via the free online configurator at www.ebi300.com
EBI 310: Data logger with internal temperature sensor
EBI 310 TE: Data logger with external precision temperature sensor, internal temperature sensor can be used additionally. Measurements in high and low temperature range
EBI 310 TH: Data logger with external humidity and temperature sensor, internal temperature sensor can be used additionally. Monitoring the relative humidity in storage and transport
EBI 310 TX: Data logger with temperature dual adapter for two exchangeable temperature sensors (not included in the scope of supply), internal temperature sensor can be used additionally
Specifications | |
Ambient temperature: Measuring range temp. EBI 310 / 310 TH: Measuring range temp. EBI 310 TE: | -30 to +75 °C -30 to +75 °C -200 to +250 °C |
Measuring range humidity EBI 310 TH: | 0 to 100 % rh |
Accuracy temperature: | ± 0.2 °C (-30 to +30 °C); ± 0.5 °C for remaining range |
Accuracy humidity EBI 310 TH: | ± 3 % (10 to 90 % rh) |
Resolution temperature / humidity: | 0.1 °C / 0.1 % rh |
Memory capacity: Sensor: | 120,000 measurements Pt1000/NTC |
Battery life time: | up to two years |
Measuring period: | 1 s to 24 h |
Maximum start delay: | 72 h |
Housing material: | Polycarbonate |
Measuring modes: | Start/Stop, endless loop, start with button |
Alarms: | 5 values |
Weight: | 35 g |
IP code: | IP 65 |
Hazardous Substance Data | |
UNNR | 3090 |
Accuracy temperature ± °C: | 0.5 |
---|---|
Depth mm: | 80 |
Description: | Temperture data loggerincl. external sensor and temp.- two-channel adapter |
Height mm: | 14 |
Measuring range temperature °C: | 75 |
Protection class: | IP65 |
Range start temperature °C: | -30 |
Resolution temperature °C: | 0.1 |
Sensor type: | Pt1000/NTC |
Storage readings: | 120000 |
Type: | EBI 310 TX |
Type of auxiliary energy: | CR 2450, 3 V |
Weight kg: | 0.064 |
Width mm: | 33 |